Original Article

RELIABILITY OF BACK CONCENTRIC ISOKINETIC MUSCLE TESTING USING BIODEX ISOKINETIC DYNAMOMETER

Volume: 19 Issue: 1, March 2004 Publish Date: March 31, 2004
Full Text PDF
Sabire Akın
Orta Doğu Teknik Üniversitesi Sağlık Merkezi FTR Ünitesi image/svg+xml
Özlem Öner
Orta Doğu Teknik Üniversitesi Sağlık Merkezi FTR Ünitesi image/svg+xml
Nisa Özberk
Orta Doğu Teknik Üniversitesi Sağlık Merkezi FTR Ünitesi image/svg+xml
Sabire Akın, Özlem Öner, & Nisa Özberk. (2004). RELIABILITY OF BACK CONCENTRIC ISOKINETIC MUSCLE TESTING USING BIODEX ISOKINETIC DYNAMOMETER. Archives of Rheumatology, 19(1), 015–019. Retrieved from https://www.archivesofrheumatology.org/index.php/pub/article/view/125

Abstract

Objective: The aim of this study is to determine the test-retest reliability of back extensor and flexor concentric muscle strength using Biodex isokinetic dynamometer.

Material and Methods: 49 healthy men and women volunteers were included. Evaluation was carried out with a Biodex System 3 Pro isokinetic dynamometer. Tests were performed in 3 sessions at a 7 day interval. Intraclass-Correlation-Coefficients were calculated to establish day-to-day repeatability. The protocol consisted of 5 isokinetic concentric contractions at a velocity of 60º/sec and 10 repetitions at a velocity of 120º/sec.

Results: Results show very good reproducibility in both movement directions at both velocities. ICC for the back flexor muscle strength at 60 º/sec were 0.88 and ICC for the extensor muscle strength were 0.84; ICC for the back flexor muscle strength at 120 º/sec were 0.85, ICC for the extensor muscle strength were 0.84.

Conclusion: The results indicate that isokinetic concentric flexor and extensor muscle strength evaluation with this test protocol is reliable. Biodex isokinetic dynamometer is a reliable tool in the evaluation of back muscular performance.

Similar Articles

Article Info
Published In
Journal Archives of Rheumatology
Volume / Issue Vol. 19 No. 1 (2004): The Archives of Rheumatology
Pages 015-019
History
Published Online March 31, 2004
License
Affiliations
1
Sabire Akın
Orta Doğu Teknik Üniversitesi Sağlık Merkezi FTR Ünitesi
2
Özlem Öner
Orta Doğu Teknik Üniversitesi Sağlık Merkezi FTR Ünitesi
3
Nisa Özberk
Orta Doğu Teknik Üniversitesi Sağlık Merkezi FTR Ünitesi
Cite this Article
Sabire Akın, Özlem Öner, & Nisa Özberk. (2004). RELIABILITY OF BACK CONCENTRIC ISOKINETIC MUSCLE TESTING USING BIODEX ISOKINETIC DYNAMOMETER. Archives of Rheumatology, 19(1), 015–019. Retrieved from https://www.archivesofrheumatology.org/index.php/pub/article/view/125
Outlines